【Scope of application】
It is used to test the tensile properties of individual fibers such as breaking strength and elongation at break.
【Relevant Standards】
【Scope of application】:It is used to test the tensile properties of individual fibers such as breaking strength and elongation at break.
[Method standard]:GB/T14337
【Technical Parameters】:
1. Load range, accuracy 0~400cN ±1cN
2. Load resolution 0.01CN
3. Elongation range 0~100mm
4. Stretching speed 5~100mm/min
5. Have a device for correct correction of the instrument;
6. The indication error of the applied force should not exceed 1%, the indication error of elongation should not exceed 0.02mm, and the error of the clamping length should not exceed 0.2mm; The instrument should maintain constant velocity elongation, and after the strength meter is started, the difference between the speed of the lower gripper and the specified speed should not exceed 5%, and the device that can make a load-elongation curve can be made.